15921758826
Shanghai Weitao automation equipment Co., Ltd
Contact person: Mr. Lu
Tel: 15921758826
Address: 2312 Weiqing East Road, Shanyang Town, Jinshan District, Shanghai
The test principle of
Under the prescribed conditions, the horizontal samples were exposed to a small ignition source, hexamethylenetetramine tablets (hereinafter referred to as tablets), and the damage length and flame spread time after the test were measured.
The instrument features
Stainless steel case with removable sliding plate.
Equipped with an observation mirror to facilitate the observation of samples.
Provide standard metal sample.
Orlotropin standard pills are available.
Stainless steel nuts are available to withstand 900℃
Test standard
Carpet flammability tester meets the following standards:
ISO 6925 ISO 6925 BS 6307-1982
CFR 1630/1631 BS 4790-1982
16.CFR 1630/1631-2012
Combine standards and user requirements.
The attachment list
When you open the package, please check the following materials:
1 test box: the internal size of the test box is 300mm×300mm×300mm. It is made of hard fire-resistant insulating board, which has the heat resistance similar to asbestos cement, and the thickness is not less than 6mm.
The top of the box is open and there is a movable bottom plate made of the same material, and the joint is sealed.
Note: Any other test box that can give the same result can be used.
Square metal plate: 230mm×230mm, thick (6.5±0.5)mm, with a circular hole with a diameter of 205mm in the middle.
Dryer: to store tablets and drying samples, it is recommended to use color-changing silica gel as a desiccant.
Oven: there should be ventilation and constant temperature control.
The temperature inside the box is (105±2)℃.
Gloves: Polyethylene, polypropylene and silica gel gloves are available.
Steel ruler: The dividing value is in millimeters.
Vacuum cleaner: All surfaces in contact with the sample should be flat and smooth.
Laboratory fume hood: the volume is about 2m3. It is airtight and the exhaust device can be closed during the test.
The front or one side of the fume hood shall be provided with a glazed window for viewing specimens.
Tablet: hexamethylenetetramine (commercial name ullotropin) flat tablet, each tablet has a mass of (150±5)mg and a diameter of 6mm.
Note: Tablets stored in a dryer can reduce the phenomenon of ignition rupture.
Timer: Stopwatch is optional.
The power supply requirements
Power supply: AC 230V, 1A, 50HZ
Fuse requirements:
Electricity inlet (F2):5AL 250V
Control circuit (F1):T500mAL 250V
Power supply system (F3):15AL 250V
To prevent fire, replace the fuse of the same type with the rated value
Test method for
2.1 Sample and humidity control
Sampling and sample selection
Intercept the sample according to QB/T 1078
Sample size and quantity
At least eight samples of 230mm×230mm and tolerance ±3mm should be taken from each sample.
Sample sampling and sample selection
Intercept the sample according to QB/T 1078
liner
The use of cushions is not specified and, if approved by the parties concerned, this method may be used to assess the combined effect of carpets and cushions.
The wet sample
Clean sample with vacuum cleaner, get rid of the loose nap on the head and other sundry.
According to the specific conditions, one of the following methods shall be used to humidify the sample. The sample shall be placed individually, with the blanket face up, or in a method agreed by the parties concerned:
Method A: According to ISO139, the sample was placed in the standard atmosphere with temperature (20±2)℃ and relative humidity (65±4)% for 24h, and then taken out and placed in a sealed container.
Method B: Put the sample in an oven at (105±2)℃ for 2h, then transfer the sample into a dryer until it reaches room temperature.
Note: The sample controlled by method B is more strict, while the sample controlled by method A is more realistic. The method should be chosen according to the specific situation.
2.2 Test Steps
1. The test should be carried out in the atmosphere with temperature ranging from 10℃ to 30℃ and relative humidity ranging from 20% to 65%.
2. Put the test chamber into the ventilator and turn off the exhaust device.
3. Put on gloves and take out a sample from the dampening atmosphere or dryer. For the pile sample, brush the pile along the direction against the pile tip to the upright position.
4. Place the sample face up on the bottom plate of the test box, place the metal plate on the sample, and align the four sides with the sample.
5. Take a tablet and place it flat on the center of the sample. Gently contact the surface of the tablet with the flame of a lit match or igniter to ignite the tablet.
If timing is required, start the stopwatch. Be careful not to touch the specimen with a lit match or the flame of the igniter.
The interval between the sample taking from the container or dryer and the tablet igniting should be controlled within 2min. If more than 2min, another sample should be taken and tested again according to 3~5.
If the tablet is broken on ignition, the test results will be invalid.
6 The ignited flame or any spreading flame shall burn until extinguished, allowing flame or non-flame to spread to any edge of the metal plate hole. If either of the above two conditions are met, the test shall be terminated, the timing shall be stopped immediately, the exhaust device of the fume hood shall be opened, and the smoke and toxic QI body shall be removed.
The maximum distance, in millimeters (mm), from the center of the specimen to the edge of the damaged area is measured with a steel ruler.
Remove the sample and remove the residue from the bottom plate of the test box.
Sufficient time shall be allowed between tests for the chamber to cool to room temperature ±5℃.
Repeat the remaining seven samples according to the method specified in 4~5.
Use a timer in seconds to measure the time it takes for the tablet to ignite into flame or non-flame to extend to the edge of the plate hole, if required.
Founded in 2011, Shanghai Huitao is a team of experienced engineers and emerging design experts. The company focuses on R & D, design and production of high quality laboratory instruments for textile testing. Headquartered in Shanghai, the company provides high-quality testing instruments and high-quality professional services for companies, academic research institutions and testing institutions of various countries.
Shanghai Huitao continuously launched products to meet the experimental needs of the textile industry. It has advanced microclimate comfort test equipment, such as thermal manikin, thermal and humidity resistance tester, MMT and other test equipment. Meanwhile, it also provides air permeability tester, hydrostatic pressure tester and thermal protective clothing performance test equipment required for new product development. In addition, we have developed and produced many basic textile testing instruments, including washing color fastness tester, fabric abrasion tester, pilling tester, rubbing color fastness tester, ironing fastness tester and perspiration color fastness tester. These high-quality products are widely used in textile testing field. In order to meet the experimental needs of users, we also provide you with innovative and practical testing instruments. The products include fiber tester, yarn tester, fabric tester, carpet tester, geotextile tester and other related testing equipment.
The company sells our products through distributed distribution partners who provide high-quality products, services and support to our customers all over the world. All the products are in line with international quality standards and specifications, and can perform outstanding and accurate product performance.